MI-ROBO

The worldwide exclusive fully automatic melt flow indexer that fills, measures and cleans itself. It enables MFR and MVR measurements for granulates and powdery materials - without any manual intervention.

  • ASTM D1238
  • ISO 1133

Customized enquiryProduct brochure


Melt flow indexer MI-ROBO

Fully automated measurements

The system automatically fills the test chamber, measures, cleans and changes the capillaries. This means that any number of measurements can be carried out in succession without the need for operator intervention.

The MI-ROBO can be customized with optional extensions to cover specific testing requirements.

Good to know

In addition to MFR/MVR measurements according to ISO 1133 procedure B and ASTM D1238 procedure B, the instrument also enables multi-weight measurements using a single barell filling to determine the flow rate ratio (FRR) according to ASTM D1238 procedure D.




Specification

Product features

  • Integrated sample magazine for 30 individual measurements
  • Fully automatic measurement with freely definable parameters for each test
  • Fully automatic cleaning of piston and test barrel with brushes and cloths
  • Integrated capillary precleaning and magazine for 30 capillaries
  • Magazine for up to eight test weights from 1 kg to 21.6 kg
  • Multi-Load measurements with two different weights
  • Force-controlled pre-compression / ejection via drive with up to 60 kg force
  • Individual filling and cleaning procedures, specific for each material
  • Electrically heated test chamber with easily exchangeable test barrel

for even more functionality

Optional units

  • Inlet design for sticky materials
  • Magazine heater
  • Nitrogen purging of sample magazine for hygroscopic polymers
  • Material feeding via conveyor system
  • Alternative steel grades for barrel and piston with higher corrosion or abrasion resistance

Best practice

Multi-Load-Measurements

Both our mi40 and MI-ROBO 89.16 can perform automated multi-load measurements with only one filling of...

 

Details

Melt Flow Indexer (MFR/MVR) vs. Capillary Rheometer

A Capillary Rheometer as well as a Melt Flow Indexer (Plastometer) can compare the flow behaviour of polymer plastics...

 

Details

Intrinsic Viscosity

The intrinsic viscosity (IV), also referred as intrinsic viscosity, or Staudinger index [η]...

 

Details

Determination of Melt Flow Index (MFR and MVR)

The melt flow index is a standardized test method to determine the flow rate of plastics at a defined temperature and constant force. The test method...

Details

Reproducible and Accuracy of Melt-Index-Tests

GÖTTFERT Melt Flow Testers (Plastometers) are manufactured according to the most modern electronic and mechanical standards with minimal manufacturing tolerances. The result is a testing device which delivers precise and accurate MVR and MFR test results.

 

Details

Better than ISO 1133

All our melt index testers are equipped with a fast and precise temperature control...